We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

We present an analysis of the effects of aberrations in third harmonic generation (THG) microscopy by considering different specimen geometries. Numerical simulations show the general trend that signal intensity and resolution are reduced as aberrations increase in amplitude. It is also shown that there are certain combinations of specimen structure and focusing position for which the presence of aberrations results in an increase in the image intensity. This occurs, for example, when there are several interfaces near the focal volume. The axial spreading of the excitation focal volume increases the characteristic coherence length for THG signal build-up, resulting in a significant contribution to the image brightness from axial planes near the focal plane. These results have important consequences for the interpretation of THG microscope images and for image based aberration correction in adaptive optics THG microscopy. © 2010 IOP Publishing Ltd.

Original publication

DOI

10.1088/2040-8978/12/8/084009

Type

Journal article

Journal

Journal of Optics

Publication Date

01/08/2010

Volume

12