We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

We examine the effects of aberrations induced by a refractive index mismatch on the signal level and resolution of single-photon (1-p) and two-photon (2-p), conventional and confocal scanning microscopes. In particular, we consider the aberrations introduced by an interface between oil/glass and water. Resolution is defined in terms of enclosed fluorescence, rather than full-width half-maximum, revealing more useful information for heavily aberrated point spread functions (PSFs). It is shown that, at large focusing depths, the resolution of 2-p conventional and 1-p confocal microscopes are almost identical. The benefits of aberration correction are examined by removing Zernike aberration modes. With aberration correction, the best resolution is found for 1-p confocal and 2-p confocal modes. An approximation based upon geometrical optics is also introduced which shows that the axial resolution of heavily aberrated PSFs is roughly proportional to focusing depth.

Original publication

DOI

10.1117/1.1382808

Type

Journal article

Journal

J Biomed Opt

Publication Date

07/2001

Volume

6

Pages

266 - 272

Keywords

Microscopy, Fluorescence, Models, Theoretical, Photons, Refractometry