We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

A computerized prediction-correction scheme has been devised for the control of alveolar gases. First, a model is run off-line to predict the inspiratory gas tensions at each second that should yield the desired alveolar patterns. Second, during the experiment, there is feedback correction based on the deviation of the actual alveolar values from the desired alveolar values. The actual alveolar values are found by a second computer and passed to the controlling computer using interrupts. The controlling computer has four digital-toi-analog outputs for controlling CO2, O2, N2, and air flows so as to achieve the commanded inspiratory PCO2 and PO2 (CO2 and O2 partial pressures, respectively). The scheme is illustrated for the generation of sinusoidal alveolar PCO2 with alveolar PO2 held constant and for steps of alveolar PCO2 at constant alveolar PO2.