Incomplete Spoiling Correction for 3D T1$$ {T}_1 $$ Mapping

Belsley G., Tyler DJ., Robson MD., Tunnicliffe EM.

Purpose: Assess the impact of incomplete spoiling and spatial saturation on (Formula presented.) mapping using the variable flip angle (VFA) spoiled gradient recalled echo sequence (SPGR). Develop a correction for incomplete spoiling that results in an accurate estimate of (Formula presented.), essential for quantitative MRI applications. Methods: A correction factor, derived from extended phase graph simulations, was applied to the measured SPGR signal to account for deviations from the theoretical steady-state signal. An iterative fitting process was used, refining the incomplete spoiling correction factor at each step based on the updated (Formula presented.) estimate. The incomplete spoiling correction method was validated against inversion recovery spin echo in phantom experiments and applied to in vivo data at 3T. Additionally, the hypothesis that spatial saturation disrupts the steady-state was also tested and validated in a phantom. Results: Without a correction, the median (Formula presented.) error was +7.4% (IQR [4.4, 9.6]%) in both simulations and phantom experiments. After applying the correction, the median (Formula presented.) error decreased to −0.3% (IQR [−2.5, 2.1]%). In vivo, incomplete spoiling caused (Formula presented.) overestimation of 10–50 ms depending on (Formula presented.) inhomogeneity. Spatial saturation induced on average a −10% (Formula presented.) bias. Conclusion: A correction for incomplete spoiling is essential to achieve an accurate measure of (Formula presented.) using the VFA SPGR method, addressing systematic (Formula presented.) overestimation when using this (Formula presented.) mapping technique. Furthermore, other mechanisms that may disrupt the steady-state signal, such as spatial saturation, should be considered.