Cookies on this website
We use cookies to ensure that we give you the best experience on our website. If you click 'Continue' we will assume that you are happy to receive all cookies and you will not see this message again. Click 'Find out more' for information on how to change your cookie settings.

Deep brain stimulation (DBS) therapy is a continually expanding field in the functional neurosurgical treatment of movement disorders. However, the occurrence of adverse events related to implanted hardware cannot be overlooked. We report on a specific feature noted in our experience of DBS-related complications. From 1998 until present we have found an overall rate of 5.3% of DBS electrode lead dysfunction (out of 133 patients) in our series (slipped leads 2.3%, lead fracture 3.8%). Interestingly, all of these failures occurred in dystonia patients (18.4% of all dystonia patients and 9.2% of all electrodes). We postulate on mechanisms that may explain why these complications predominate in this group of patients.

Original publication




Journal article


J Clin Neurosci

Publication Date





243 - 245


Adolescent, Adult, Aged, Aged, 80 and over, Child, Deep Brain Stimulation, Dystonia, Electrodes, Equipment Failure, Female, Foreign-Body Migration, Humans, Male, Middle Aged, Movement Disorders, Neurosurgical Procedures, Retrospective Studies