We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

We describe the theory of a new method of optical refocusing that is particularly relevant for confocal and multiphoton microscopy systems. This method avoids the spherical aberration that is common to other optical refocusing systems. We show that aberration-free refocusing can be achieved over an axial scan range of 70 μm for a 1.4 NA objective lens. As refocusing is implemented remotely from the specimen, this method enables high axial scan speeds without mechanical interference between the objective lens and the specimen. © 2007 Elsevier B.V. All rights reserved.